Reference pressure transmitters ElMetro-Pascal-04
The principle of operation of the transducers is based on the use of the relationship between the measured pressure and the elastic deformation of the membrane of the sensing element. The measured pressure acts on the sensitive element of the transducer, which leads to the appearance of an electrical signal at its output. The electronic module of the converter amplifies the electrical signal and converts it, taking into account the temperature of the sensing element, into a measured value, available as a digital output signal.
The transducers consist of a sensing element and an electronic module housed in a metal case.
The converters have an open digital output communication protocol. To power the transducers and read the measured pressure value from the digital output of the transducer, the following are used:
- multifunctional calibrators ELMETRO-Pascal-03 (registration number in the federal information fund 73828-19), other similar devices with a compatible digital interface;
- a personal computer equipped with a USB port - together with a USB adapter and software (hereinafter referred to as software) "ARM-Pascal" or other software that provides interaction with converters.
Multifunctional calibrators ELMETRO-Pascal-03 do not introduce additional error when displaying the measured pressure value.
By design, the converters have versions that differ in the size of the housing, the number of pressure connection inputs and the number of interfaces.
The general view of the converters is shown in Figure 1. The scheme of sealing the converters against unauthorized access is shown in Figure 2. The sealing is carried out by the manufacturer.
Transmitter designation structure:
ELMETRO-Pascal-04
-
Х
-
Х
-
Х
-
Х
1
2
3
4
5
1 - Name of the converter (ELMETRO-Pascal-04);
2 - Code of execution of the converter for explosion protection (Ex or empty field);
3 - Modification of the converter in accordance with table 1;
4 - Accuracy class or number equal to the limit of the permissible basic absolute error in accordance with table 1;
5 - Temperature range code (T35, T50) in accordance with table 2;
Table 1 - Metrological characteristics of converters of absolute, gauge pressure, pressure-vacuum
Performances
Subrange 11)
Subrange 21)
Accuracy (α)
αadd5)
Range (from LLM2) to
ULM3)), MPa
Transient pressure
(PП), MPa
Range (from LLM2) to
ULM3)), MPa
Transient pressure
(PП), MPa
relative |P4)| ≥|PП|, %
reference PП, | P4) | <|PП|, %
Overpressure, underpressure
1K
from 0 to 0,001
0,001
–
–
±0,05; ±0,1; ±0,15; ±0,2
0,5
1KР
from -0,001 to +0,001
-0,001;
+0,001
7K
from 0 to 0,007
0,004
from 0 to 0,004
0,0016
±0,04; ±0,05; ±0,06
7KР
from -0,007 to +0,007
-0,004;
+0,004
from -0,004 to +0,004
-0,0016;
+0,0016
40K
from 0 to 0,04
0,02
from 0 to 0,02
0,01
±0,02; ±0,025; ±0,03; ±0,04; ±0,05
40KР
from -0,04 to +0,04
-0,02;
+0,02
from -0,02 to +0,02
-0,01;
0,01
160K
from 0 to 0,16
0,08
from 0 to 0,08
0,04
±0,01; ±0,02; ±0,025; ±0,03; ± 0,04; ±0,05
160KР
from -0,1 to +0,16
-0,08;
+0,08
from -0,08 to +0,08
-0,04;
+0,04
1M
from 0 to 1
0,5
from 0 to 0,5
0,25
1MP
from -0,1 to +1
from -0,1 to +0,5
7М
from 0 to 7
3,5
from 0 to 3,5
1,6
±0,02; ±0,025; ±0,03; ±0,04; ±0,05
7МP
from -0,1 to +7
from -0,1 to +3,5
25М
from 0 to 25
16
from 0 to 16
10
60М
from 0 to 60
35
–
–
Absolute pressure
160KA
from 0 to 0,16
0,1
–
–
±0,02; ±0,025; ±0,03; ±0,04; ±0,05; ±0,06
0,5
1MA
from 0 to 1
0,5
from 0 to 0,5
0,25
1) The transducers have the ability to programmatically switch measurement subranges by the user;
2) LLM - the lower limit of measurements;
3) ULM - upper limit of measurements;
4) P is the value of the measured pressure;
5) αadd- The limits of the permissible additional relative or reduced to Pp error caused by a change in the ambient temperature in the operating temperature range for every 10 ºС, in fractions of the limits of the main relative or reduced to P measured pressure).
Table 2 - Main technical characteristics of converters
Description of characteristics
Value
Digital interface
UART
Normal measurement conditions:
- ambient temperature, ºC
- relative air humidity,%, no more
- atmospheric pressure, kPa
from +10 to +30
80
from 84 to 106,7
Measurement working conditions:
- ambient temperature, ºC
- for code T35
- for code T50
- relative humidity at a temperature of +35 ºC,%, no more
- atmospheric pressure, kPa
from +5 to +35
from -10 to +50
90
from 84 to 106,7
Weight, kg, max
1,5
Overall dimensions, mm, no more:
- design 1 (length × body diameter)
- design 2 (length × body diameter)
- design 3 (length × height × width)
98x41
123x41
109x67x72
Mean time between failures, h
30000
Average service life, years
8